NORMAL START UP SEQUENCE
EVERY TIME YOU USE THE ANALYZER
BEFORE SWITCH-ON CHECK THAT:
• the oxygen sensor is connected
• the particle filter is not dirty
• the sulphur filter is fitted for heavy oil or coal
• the water trap and probe line are empty of water
• all hose connections, etc, are properly made
• the paper roll is fitted
• the analyzer unit is in fresh air
• the water trap is vertical
• the flue temperature is connected
• the instrument is placed on a clean, flat, level surface
Switch ON the analyzer by pressing ON/OFF on the handset.
You also need to press the ON/OFF
switch on the analyzer main unit.
AUTOMATIC CALIBRATION
• During this sequence the analyzer pumps fresh air into the sensors to allow toxic sensors to be set to
zero and the oxygen sensor to be set to 20.9 %.
• During this sequence the handset display will show the following: