32
7.18
Replacing the water pump
Caution!
Before starting, disconnect the instrument from the power source.
1. Remove the autoclave cover (see para. 9.1
“
Dismantling the Outer
Covers of the Autoclave
”
).
2. Disconnect the wires from the pump
3. Empty the water reservoir.
4. Disconnect the piping from the pump.
5. Remove the pump from the rubber shock absorbers.
5.1 If the rubber shock absorbers are damaged, replace them. Note
that on models 2340/2540 the shock absorber’s bases are
pointed outward and on models 3140/3850/3870 they are
pointed inward.
6. Replace the damaged pump with a new pump.
7. Reconnect wiring and piping.
8. Reassemble the cover.
9. Turn on the autoclave and verify it operates correctly.
No.
Description
2340/2540
3140/3850/3870
1
Pump, Water, EXA7, 220V, Ulka PUM055-0014 PUM055-0014
2
Rubber shock absorber
SKR203-0006 SKR203-0006
3
Hose adaptor, male, elbow,
1/8BSP, for 6 mm Hose.
FIT100-0806 FIT100-0806
4 Screw
BOL191-0140 BOL194-0342
5
Nut, Hex, Flange, 1/4NC
NUT192-0155
6 washer
NUT193-0347 NUT193-0276
7
Rivet, Dome Head, Aluminum,
4x14
BOL194-0331
2340/2540
3140/3850/3870
1
2
6
4
5
1
2
6
7
3
3
Summary of Contents for 3850 ELVG-D
Page 43: ...42 AJUNC 3 BOARD...
Page 53: ...52 VESSEL ASSEMBLY...
Page 55: ...54 OUTER CABINET ASSEMBLY...
Page 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Page 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Page 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Page 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Page 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Page 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...