26
7.13
Replacing the Printer
7.13.1 DPU20
Printer
Caution!
Before starting, disconnect the instrument from the power
source.
1.
Remove the autoclave cover (see para. 7.5
“
Dismantling
the Outer Covers of the Autoclave
”
).
2.
Disconnect the connector of the flat cable (1) connecting
the electronic box to the printer (2).
3.
Unscrew the two screws (3) attaching the fastening
bracket (4) to the printer.
4.
Remove the printer and insert the new one.
5.
Assemble the fastening bracket to the printer with the
two screws (2) and verify that it "sits" firmly in its seat.
6.
Connect the flat cable (1) to the printer.
7.
Verify connection of power by performing a self-test.
8.
Run a cycle and verify that the printer operates correctly.
1
3
2
4
No.
Cat. No.
Description
1 WIR040-0070 Cable, Flat, Printer, DPU20, 25CM, 34P
2 THE002-0005 Printer,
DPU-20, Seiko
3
BOL190-0144 Screw, printer tightening
4
Supplied with
the printer
Fastening bracket
Summary of Contents for 3850 ELVG-D
Page 43: ...42 AJUNC 3 BOARD...
Page 53: ...52 VESSEL ASSEMBLY...
Page 55: ...54 OUTER CABINET ASSEMBLY...
Page 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Page 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Page 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Page 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Page 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Page 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...