37
8.1
Changing the parameters
Not all the switches are relevant to this autoclave, details will be given
for the relevant switches only:
Note: Changes must be done when the autoclave is shut off.
SWITCHES
1,2
Selecting an identification number for the autoclave
1 ON,
2 ON = number 1
1 ON,
2 OFF = number 2
1 OFF, 2 ON = number 3
1 OFF, 2 OFF = number 4
SWITCH
3
Change of parameters
ON
= Unable to change option
OFF
= Able to change option
SWITCH
4
Stand by shoot mode
ON
=
Able
Off = Unable
SWITCHES
5,6
Selection of temperature and pressure parameters
5 OFF, 6 OFF = °F ,psi
5 OFF, 6 ON = °F ,kPa
5 ON,
6 OFF = °C ,psi
5 ON,
6 ON = °C ,kPa
SWITCH
7
Factory use only.
SWITCH
8
Option for connecting a printer.
ON
printer is connecting.
OFF
printer not connected.