29
7.15
Cleaning water inlet strainer
Caution!
Before proceeding, make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
1. Remove the cover of the autoclave.
2. Drain the water from the water reservoir.
3. Remove the water filter from the silicon tube.
4. Open the filter by unscrewing the two filter parts.
5. Clean the filter by flushing it under running water for a few
minutes.
6. Replace the filter parts and reconnect it to the silicon tube.
7. Open the water valve.
Water Reservoir Filter