10
6.9
Inlet key does not let water into the chamber
♦
Check if this key is faulty. If it beeps when pressed it is O.K. if not,
replace the keypad. If this does not help, replace the complete
PREDG and keypad unit.
♦
Check the float at the water reservoir by manipulating it using a
screwdriver or a tool and repair/replace it. If the float is
functioning, the problem could be with the control circuit at the
PREDG board: the test board between TP1 (GND) and TP10 (water
valve): GOOD = 0.0V to 0.05V.
♦
If the control is OK, there may be a problem with the valve itself or
with its connector.
♦
Check and replace the coil if necessary.
6.10 Pressure stays close to 100 kPa (0 psi), temperature does not exceed
110 - 115
°
C (207
°
F) or, water inlet lasts a longer time than usual
This phenomenon is most likely due to malfunction of the air valve. The
air valve’s normal operation is as follows:
OPEN whilst turning the system on, OPEN throughout WATER INLET
and HEAT until temperature has reached 90
°
C or 194
°
F, CLOSED from
that stage on, and OPEN during DRY.
The valve’s control on the AJUNC3 board should act as follows:
0.0V = valve OPEN
5.0V = valve CLOSED
If the problem lies on the AJUNC 3 board it should be replaced. It is
recommended to replace both the AJUNC3 and the solenoid because a
damaged solenoid may affect the AJUNC3 board. If the control circuit
on the AJUNC3 is OK, the problem is probably with the PREDG board.
6.11 Heaters are always ON or OFF
Warning: If you need to tighten a screw, make sure to do so only when
the power cord is disconnected.
♦
Check the heaters connections.
♦
Check
the
cutoff
thermostat
position
and
turn
it
clockwise
to
the
end.
♦
Check the control at TP1 and TP12 for ON = 0.0V, OFF = 5v
♦
Check connections between AJUNC3 to SSR at TP4.
♦
Does AC voltage enter the SSR at AC connector and SSR pin 2?
♦
Is
SSR
output
the
AC
voltage?
Check
AC
connector
and
at
SSR
pin
1.
♦
Is the control OK? Check the SSR for 5.0V between pins 3&4. If
control
is
OK,
but
SSR
output
is
not
the
AC voltage, replace the
SSR.
6.12 Heaters o.k. but autoclave does not reach sterilization temperature
♦
Check and calibrate the cutoff position.
♦
Check and clean/ replace leakage at the exhaust valve.
♦
It is also possible that there is a calibration variance, check the
GAIN pressure calibration, perhaps it should be lowered or, the
GAIN temperature should be increased.
♦
Another possibility is that insufficient water level in the chamber
does not allow the pressure to build up. Check the water level
detector electrode, or the leveling of the autoclave.
Summary of Contents for 3850 ELVG-D
Page 43: ...42 AJUNC 3 BOARD...
Page 53: ...52 VESSEL ASSEMBLY...
Page 55: ...54 OUTER CABINET ASSEMBLY...
Page 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Page 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Page 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Page 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Page 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Page 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...