REPLACEMENT OF COMPONENTS
7.1
Safety Tests after Repair
ATTENTION!
After every repair or dismantling the enclosure, the autoclave
should pass two safety electrical test by the Service Engineer. The
following shall be performed:
1. Enclosure Leakage Current Test.
Every autoclave should pass this test as follows:
1. Connect the electrical cord to the autoclave.
2. Turn on the main switch and the circuit breaker.
3. Short-circuit the L and N pins on the cord's plug.
4. Connect the Short-circuit pins to the L pole on the Megger.
5. Connect the earth pins to the earth pole on the Megger.
6. Impose an electrical potential of 500-1000V on the tested
autoclave. The insulation resistance should be at least 2 M
Ω
.
The test is successful if there was no leakage.
2. Protective Earth Impedance Test
1. Connect the grounding pin of the power cord plug to one pole of
an Ohmmeter.
2. Connect any other metallic part (preferable – the metallic part of
the locking screw) to the second pole of the Ohmmeter.
3. The resistance should not exceed 0.3
Ω
.
After performing these tests, the Service Engineer should complete and
sign the Work Order.
Summary of Contents for 3850 ELVG-D
Page 43: ...42 AJUNC 3 BOARD...
Page 53: ...52 VESSEL ASSEMBLY...
Page 55: ...54 OUTER CABINET ASSEMBLY...
Page 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Page 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Page 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Page 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Page 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Page 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...