20
7.7
Replacing the Temperature Safety Thermostat
The autoclave is supplied with a temperature thermostat, which protects
the heaters and autoclave against overheating, during the dry cycle.
This device reconnects automatically when the chamber cools down.
Caution
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing outer covers.
The temperature safety thermostat is located on the lower side of the
fuse and socket panel on the rear of the autoclave.
1. Remove the rear cover.
2. Loosen the heating band.
3. Unscrew the thermostat and replace it with a new one.
4. Perform any dry cycle to verify that the
temperature safety
thermostat disconnects the heating units.
Safety Thermostat
Summary of Contents for 3850 ELVG-D
Page 43: ...42 AJUNC 3 BOARD...
Page 53: ...52 VESSEL ASSEMBLY...
Page 55: ...54 OUTER CABINET ASSEMBLY...
Page 59: ...58 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 60: ...59 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 61: ...60 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 62: ...61 ELECTRICAL DIAGRAM FOR MODELS 1730 E EK...
Page 63: ...62 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 E EK...
Page 64: ...63 ELECTRICAL DIAGRAM FOR MODELS 2340 2540 EA EKA...
Page 65: ...64 ELECTRICAL DIAGRAM FOR MODEL 3140 E...
Page 66: ...65 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 E EK...
Page 67: ...66 ELECTRICAL DIAGRAM FOR MODELS 3850 3870 EA EKA...