Zero-Gravity® Radiation Protection System Uncrating Instructions
29
Floor & Ceiling Mounted Units
ZGM-6-5H | ZGHSA | ZGCM-HSA | ZGCM-48 | ZGCM-66
Unpack the Body Shield Assembly
CAUTION: HEAVY EQUIPMENT
Use care when handling leaded body shield assembly. The assembly
weighs ~ 54 pounds (24.5 kilograms) and must be handled by two people.
WARNING! RADIATION DAMAGE HAZARD
The Body Shield must be handled with care to avoid damage to the Body
Shield protective lead material. If the Body Shield is damaged, it must be
inspected according to the Body and Face Shield Fluoroscopy Inspection
section of the User’s Guide (TIDI Products document 84000).
1. Remove the screws that secure the Straps (shown in red) to the Body Shield.
2. Leaving the protective foam cover intact (Figure 19), remove the body shield
assembly from the crate and place on a clean, flat surface Protective covers
remain until installation is complete.
Figure 19
Screws
Straps
Protective Foam Cover