![Tianlong Gentier 48E User Manual Download Page 23](http://html.mh-extra.com/html/tianlong/gentier-48e/gentier-48e_user-manual_1114100023.webp)
9
Gentier Real-Time PCR System
2. Working Environmental Requirements
The Gentier instrument is for indoor use only.
2.1 Instrument Space Requirements
1)
The Gentier instrument should be placed in the room with low humidity (between 20%
~85%RH) and appropriate temperature (between 10°C~30°C). The room should be well
ventilated and without corrosive gas.
2)
The Gentier instrument should be placed on steady lab work bench and leave enough space
for placing the attachments, cables and reagent bottles, etc.
3)
Please keep the Gentier instrument away from heat sources (directly sunshine, heater, stoves)
and water sources (such as water pool, water tube etc.).
4)
The working environment of Gentier instrument should be without electromagnetic
interference, vibration and high frequency wave electrical equipment.
5)
For single Gentier instrument running, the distance between adjacent objects and the
surrounding ventilation openings shall be no less than 30cm.
6)
For protection against overheating hazards, the openings on the Gentier instrument are
designed for ventilation. Please do not block or cover these ventilation openings while it is
running.
7)
Do not place the Gentier instrument on soft surface, the instrument base may sink into the
soft surface and block up the air inlet beneath.
8)
Do not place the Gentier instrument close to the wall or pile up other goods at the rear of the
instrument, this may affect heat dissipation.
2.2 Instrument Power Requirements
1)
The power grid environment of Gentier instrument should possess ground wire and the
instrument should be properly grounded.
2)
Please ensure the power strip of Gentier instrument has 3~4 three-phase plugs in order to
meet the demands for the instrument, control computer and printer.
Summary of Contents for Gentier 48E