Gentier Real-Time PCR System
Figure E-3. Run Setting interface - Temperature Setting sub-tab
2.2.1 Temperature Setting
The
Temperature Setting
sub-tab is selected on the run setting interface by default, as shown in
figure E-3. User can edit the relevant experiment settings and temperature program in the
Temperature Setting
sub-tab.
Interface Introduction &
1.
Tube Type
: the instrument software provides three tube type opetions for user, including
Clear
,
White
and
Frosted
. User can press
icon to select.
2.
Reaction Volume
: user can press
or
icons to set the reaction volume.
Reminding
: the reaction volume setting range is 0µL~100µL.
3.
Hot Lid
: user could press the check box of
Hot Lid
to decide whether to utilize the hot lid
heating function, and press
or
icons to edit the hot lid temperature.
Reminding
: the hot lid temperature setting range is 40°C ~110°C.
4.
Stage Setting Box
:
Stage
: display the stage type of the temperature program.
Cycle
: display the cycle number of the current stage.