97
Gentier Real-Time PCR System
▶
Seven predefined stage types are provided by the instrument software for user:
Preincubation
,
Reverse Transcription,
2 Step Amplification
,
3 Step Amplification
,
Melting
,
Continuous Melting
and
Cooling
. User can also press
Custom Stage
to define the
stage setting according to specific experiment requirements.
▶
User can press
< Add >
to add the selected stage to stage setting box.
▶
User can press
< Back >
to return back to the run setting interface.
Reminding
: at least one stage should be included in the temperature program.
Step 2
:
Edit stage
▶
The added stage will be automatically displayed in the stage setting box and user can select
any stage to delete or change its sequence.
▶
User can select a certain stage and press its cycle number in the corresponding
Cycle
column,
the cycle edit box will pop up as shown in figure E-3.2.
Figure E-3.2 Cycle edit box
User could set the cycle number in the cycle edit box, press
< OK >
to confirm the cycle setting
or press
< Cancel >
to cancel the setting.
Reminding
: the cycle setting range is 1~99.
Step 3
:
Add/delete a stage
▶
Select a stage in the stage setting box and the step setting box on the right will display the
corresponding steps of the selected stage;
▶
User can add a new step or select a step to delete or change its order;
Reminding
: at least one step should be included in one stage.