Thermo Scientific
49iQPS Instruction Manual
1-1
Chapter 1
Introduction
The Thermo Scientific™ 49iQ Ozone Primary Standard utilizes UV
Photometric technology to reliably produce ozone up to 1000 ppb in
concentration.
The 49iQ Ozone Primary Standard is a dual cell photometer, the concept
adopted by NIST for the national ozone standard. The instrument can
accurately generate up to 1000 ppb of ozone operating with ozonator flow
rates between 3-4 liters per minute. Because the instrument has both
sample and reference flowing simultaneously, a response time of 20 seconds
can be achieved. Temperature and pressure correction are standard
offerings.
The iQ Series Instrument Platform is a smart environmental monitoring
solution for ambient and source gas analysis that affords greater control
over instrument performance and data availability.
●
Distributed Measurement and Control (DMC) module design
simplifies serviceability. Each DMC module contains its own
microprocessor control enabling functional performance validation
at the module level.
●
Built-in predictive diagnostics and preventive maintenance
schedules identify problems before they occur. The iQ Series
platform sends email notifications directly to Thermo Fisher
Scientific’s world class service support team or locally identified
addressees in order to proactively communicate instrument
performance conditions and identify spare parts needs before an
operational concern arises.
●
The iQ Series platform supports Modbus, streaming and VNC
protocols over serial and Ethernet as well as analog and digital I/O
for easy integration into most data management systems.
●
Three standard USB ports afford convenient data download
capability as well as the ability to connect additional hardware, such
as a computer keyboard or mouse.
●
The iQ Series GUI runs on a 7” color touchscreen display. The
GUI is highly flexible and can be customized to enable a tailored
iQ Series
Instrument
Platform
Summary of Contents for Ozone Primary Standard 49iQPS
Page 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Page 2: ......
Page 8: ......
Page 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Page 142: ......
Page 208: ......
Page 216: ......
Page 234: ......
Page 242: ......
Page 254: ......
Page 255: ......