![ThermoFisher Scientific Ozone Primary Standard 49iQPS Instruction Manual Download Page 168](http://html1.mh-extra.com/html/thermofisher-scientific/ozone-primary-standard-49iqps/ozone-primary-standard-49iqps_instruction-manual_1105615168.webp)
Servicing
LCD Module Replacement
6-12
49iQPS Instruction Manual
Thermo Scientific
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1.
Turn instrument OFF and unplug the power cord.
2.
Gripping from the top corners of the front panel and pull outwards.
3.
Unfasten four nuts (
Figure 6–13
).
Figure 6–13.
Replacing the LCD Module
4.
Remove cover.
5.
Unplug LCD cables from backside of board.
6.
Pull board off the standoffs.
LCD Module
Replacement