Optional Equipment
Zero Air Source
Thermo Scientific
49iQPS Instruction Manual
8-17
The zero air source option is a convenient system for generating pollutant
free zero gas for O
3
monitoring requirements. This option includes the
pump, filter, filter holder, and Silastic tubing.
A 5-10 micron pore size, two-inch diameter PTFE element is available for
the 49iQPS. This filter should be installed just prior to the SAMPLE
bulkhead. When using a filter, all calibrations and span checks must be
performed through the filter.
Zero Air Source
PTFE Particulate
Filter