![ThermoFisher Scientific Ozone Primary Standard 49iQPS Instruction Manual Download Page 222](http://html1.mh-extra.com/html/thermofisher-scientific/ozone-primary-standard-49iqps/ozone-primary-standard-49iqps_instruction-manual_1105615222.webp)
Optional Equipment
Connecting External Devices
8-6
49iQPS Instruction Manual
Thermo Scientific
Table 8–4.
Analog Voltage and Current Assignment
Terminal Number
Signal Name
1
Current Out 1
2
Voltage Out 1
3 C/V
Return
1
4
Current Out 2
5
Voltage Out 2
6 C/V
Return
2
7
Current Out 3
8
Voltage Out 3
9 C/V
Return
3
10
Current Out 4
11
Voltage Out 4
12 C/V
Return
4
13
Current Out 5
14
Voltage Out 5
15 C/V
Return
5
16
Current Out 6
17
Voltage Out 6
18 C/V
Return
6
The iQ series instruments provide for the ability to calibrate the analog
outputs (both voltage and current) of the instruments. The basic procedure
for both voltage and current are the same using the following procedure:
●
Complete the connections of the recording device to the desired analog
output channel. (See page
8-5
for the channel information).
●
Calibrate the output channel low level.
Note
When calibrating the current output when using the 0-20 mA
scale, the low level will be set to 4 mA due to the inability to adjust the
actual current output to below zero.
▲
●
Calibrate the output channel full scale.
Analog Output
Calibration
Summary of Contents for Ozone Primary Standard 49iQPS
Page 1: ...49iQPS Instruction Manual Ozone Primary Standard 117434 00 1Apr2020...
Page 2: ......
Page 8: ......
Page 31: ...Operation Instrument Display Thermo Scientific 49iQPS Instruction Manual 3 7...
Page 142: ......
Page 208: ......
Page 216: ......
Page 234: ......
Page 242: ......
Page 254: ......
Page 255: ......