Servicing
49iQPS Photometer DMC
6-30
49iQPS Instruction Manual
Thermo Scientific
Use the following procedure to remove and replace the photometer lamp.
Equipment required:
Hex drive, 3/32
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Disconnect lamp connector.
3.
Unhook lamp cable from cable holder.
Figure 6–35.
Lamp Removal pt 1
4.
Unfasten Velcro strap.
Figure 6–36.
Lamp Removal pt 2
Photometer Lamp
Replacement