Servicing
49iQPS Photometer DMC
Thermo Scientific
49iQPS Instruction Manual
6-37
Use the following procedure to replace the thermistor.
Equipment required:
Wrench
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug the thermistor from the photometer DMC board.
3.
Using a wrench, unfasten 1/4 hex nut.
4.
Pull outward and replace.
5.
Reassemble in reverse order.
Figure 6–43.
Replacing the Thermistor
Thermistor
Replacement