Operation
Settings
3-92
49iQPS Instruction Manual
Thermo Scientific
Home Screen>Settings>Instrument Settings>Advanced Instrument
Settings>49iQPS Plumbing>Update Plumbing
The Update Plumbing screen contains the following information:
●
One Sample Capillary:
If the instrument has 3 total capillaries (1
Sample capillary, 1 Reference Flow capillary, 1 Ozonator capillary) and
a Flow/Pressure DMC that has 2 absolute pressure sensors on it, the
user would choose One Sample Capillary.
●
Two Sample Capillary:
If the instrument has 4 total capillaries (2
Sample capillaries, 1 Reference Flow capillary, 1 Ozonator capillary)
and a Flow/Pressure DMC that has 3 absolute pressure sensors on it,
the user would choose Two Sample Capillaries.
Note
After selecting the number of capillaries, the user will be required to
reboot the instrument for the changes to the user interface to take effect.
▲