Herasafe 2025
50161258 |
45
Operation |
Chapter 6
Audible Warn Signal
When the front window is moved out of the working position or when the pressure sensors
detect a safety-relevant change of the airflow velocities, the corresponding visual and
audible alarm signals are issued. The alarm remains active until the front window has
been moved to the correct working position or until the airflow velocity corresponds with
the preset values.
Switching the Illumination ON and OFF
In each operating mode, excluded UV-disinfektion,the sample chamber illumination can
be switched on or off.
Switching illumination on or off:
Shortly press key
Activating/Deactivating Internal Power Supply
All sockets in the sample room are supplied with power in a switching operation or
de-energized.
1. Switch on power supply:
Shortly press key
The blue status indicator
is illuminated.
2.
Switch off power supply:
Shortly press key
The status indicator
goes off.
Changing UV-Desinfection Time
This value refers to the set run time of the UV disinfection (optional). This display function
is only available when the front window is not closed. The device must be switched to
work mode (see Setting UV-Disinfektion).
Unsafe working!
When the alarm signals are activated, safe working is no longer ensured!
NOTE Switching the alarm signals off.
If the alarm signals fail to be switched off automatically, move the front
window to the lowest position and contact the Technical Service.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...