| Operation
42
|
50161258
Herasafe 2025
Chapter 6
If the internal power supply is ON, the blue status indicator INTERNAL POWER
SUPPLY ACTIVATED is illuminated.
The monitor contact for the external device connection is ready for operation:
The switching state of the potential-free contact is switched through (for example, if
the solenoid valve of the gas supply line is activated via this contact, gas is supplied
only in this mode).
The display can show values for: Operating hours, downflow velocity, and exhaust
airflow velocity or the time, according to the selection of the operator.
Window-Open-Mode:
Designates the condition when the window is open and
located outside the working position. To install and remove accessories, the front
window can be opened or moved upwards.
The airflow system blowers are switched on :
The red status indicator AIRFLOW NOT STEADY Is illuminated.
The front window is not in the working position:
The red status indicator FRONT WINDOW NOT IN WORKING POSITION is
illuminated.
The work space illumination is operational.
The audible alarm is active.
The internal power supply is operational:
If the internal power supply is switched on, the status indicator INTERNAL
POWER SUPPLY ACTIVATED illuminates.
Stand-By-Mode:
For an interruption of the work process, the front window can be
lowered and the sample chamber sealed aerosol-tight. The air system output has been
reduced to match the lower air requirement.
The front window is closed:
The air system operates at reduced output.
The red status indicator FRONT WINDOW IS NOT IN WORKING POSITION is
illuminated.
The blue status indicator AIRFLOW REDUCED is illuminated.
The sample chamber illumination is available.
The internal power supply in the sample chamber is available:
If the internal power supply is ON, the blue status indicator INTERNAL POWER
SUPPLY ACTIVATED is illuminated.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...