![ThermoFisher Scientific HERASAFE 2025 Operating Instructions Manual Download Page 22](http://html.mh-extra.com/html/thermofisher-scientific/herasafe-2025/herasafe-2025_operating-instructions-manual_1105592022.webp)
| Unit Description
22
|
50161258
HERAsafe 2025
Chapter 4
Test hoses for the exhaust air unit [9] and the downflow unit [13] are under the work
plates.
Optional stands [12], adjustable in height and with fixed height.
Side slates with two sealed access openings [14]. These can be equipped with media
valves.
Internal outlets [15] for power supply of accessories.
Lighting device [16], equipped with 2 tubes, is mounted on the front window.
Workplate segments [10] with arm rests [11]. A one-piece workplate and special
work plates are available options..
Safety System
The safety system comprises a combination of protective and alarm systems that ensure
maximum personal and material protection.
Safety systems:
Vacuum-sealed air system
An air system in combination with HEPA filters for downflow and exhaust air forms
the basis of the safety system for personal and material protection.
Personal protection
Air aspired from the exterior along the entire working opening at a constant high
speed prevents that:
agents may leak through the working opening of the chamber.
As the exterior air pressure around the unit exceeds the pressure of the internal air
system, it is ensured that:
agents cannot be released to the exterior in the case of a leak in the cabinet
housing.
Product protection
A steady airflow within the air system ensures that:
a constant downflow allows the HEPA filters to remove contaminants so that the
samples are always surrounded by ultrapure air,
harmful particles are not carried over through the sample chamber (protection
from cross-contamination).
NOTE Test hoses
Do not remove the two test hoses for checking downflow and exhaust air.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...