HERAsafe 2025
50161258 |
27
Unit Description |
Chapter 4
Figure 7
Access through Front Window
Front window:
The electrically operated front window [1] is made of multi-layer safety glass and integral
to the front cover frame. It can be raised to a maximum opening height B. To access the
sample chamber during the work process, the front window must stay in the work position
with opening height A.
Lowering the front window when the device is deenergized:
The safety feature is backed up by a battery. Should a power failure occur, the auxiliary
rocker switch (see “Using the auxiliary Rocker Switch” on page 52 ) can be used to lower
the front window completely.
A - minimum working opening
B - maximum opening height
Crushing hazard
During movement of the front window do not try to access the inner chamber.
Front window movement!
Do not attempt to move the front window manually as otherwise the motor drive
may be damaged.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...