| Start-up
36
|
50161258
HERAsafe 2025
Chapter 5
Power Supply Connection
Establishing the power supply connection:
1. Before connecting the device to the power supply system, check to see if the voltage
of the outlet corresponds with the specifications on the nameplate of the device. If
the ratings given for voltage (V) and maximum current (A) are not correct, the device
must not be connected to the power supply system.
2. Connect the grounding plug of the device to a properly grounded and fused outlet.
The outlet must be fused separately using a fusible link T 16 A or using a circuit
breaker B 16.
3. Make sure that the power supply line is not subjected to tensile or compressive force.
Installation of the power supply connection:
To protect against accidental switch-off, the outlets for the connection to the power supply
must be located outside the normal hand reaching range and must be accessible only to
authorized personnel. Ideally, the outlets should be installed above the safety cabinet and
be readily accessible. Route the power cables so they do not run across hot surfaces
(such as exhaust tubing or similar tubing runs).
Connecting the equipotential bonding:
If the sample chamber is supplied with media (gas, water, etc.), the on-site equipotential
bonding must be connected to one of the premounted threaded bushings at the top of the
housing.
Initialization routine:
After the unit has been connected to the power supply system, the device control runs
through a start-up initialization routine.
USB-Port
Figure 13
USB-Port
The USB-Port on the top of the device is reserved for service work!
High voltage!
Contact with current-carrying components may cause a lethal electric
shock.
Before connecting the device to the power supply system, check plug
and power supply cable for possible damage. Do not use damaged
components to connect the device to the power supply system!
The connection has to be carried out according to country specific
standards.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...