| Cleaning and Decontamination
58
|
50161258
Herasafe 2025
Chapter 9
Recommended disinfectants:
Predisinfection:
1. Remove all samples from the sample chamber and store them properly.
2. Remove accessories from the safety cabinet and disinfect them using the disinfection
procedure recommended by the manufacturer.
3. The workplate and stainless steel components can be removed from the sample
chamber and disinfected separately.
4. For predisinfection, spray disinfectant on all sample chamber surfaces or wipe the
surfaces using disinfectant.
5. Do not remove the optional UV lamps from the sockets; wipe them thoroughly using a
damp cloth.
6. Switch the device to work mode, move front window to the working position.
7. Allow disinfectant to react as recommended by manufacturer, then operate the safety
cabinet for at least 15 to 20 minutes in the work mode so that released aerosols can
be absorbed by the filters.
Cleaning:
1. Thoroughly remove dirt residues and deposits using a solution of tepid water and
dishwashing agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the drain pan and wipe all sample chamber surfaces dry.
Final disinfection:
1. Again, spray disinfectant on all sample chamber surfaces or wipe the surfaces clean
with disinfectant.
2. Allow disinfectant to react as recommended by manufacturer.
NOTE Compatibility
Chloride-containing disinfectants may damage some surfaces; use only
chloride-free disinfectants or a disinfectant with a low enough chloride content
to have been proved harmless for stainless steel finishes! Disinfectants with
an alcohol content of more than 70 % may cause embrittlement of plastic
components after extended exposure. Use only disinfectants with a low alcohol
content. When using a disinfectant with an alcohol content of more than 70 %,
the release limit of 200 g within 2 hours must not be exceeded. Also suited are
disinfectants based on quaternary ammonium compounds.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...