Herasafe 2025
50161258 |
43
Operation |
Chapter 6
UV-Mode:
For running the UV disinfection routine, the front window is completely lowered
to protect against UV radiation. The routine cannot be run until the front window is in the
"closed" position.
The UV-Disinfections routine is aktive:
The yellow status indicator UV DISINFECTION ROUTINE ACTIVATED is illuminated
until the preset time for the routine has elapsed. Then, the UV lamps are
switched off automatically, and the status indicator is switched off.
The sample chamber illumination is available.
The internal power supply in the sample chamber is not available.
The integral UV lamp (optional) is available.
The 7-segment display shows the remaining time until the complete UV routine has
expired.
Operation
The device control software automatically determines the temporary operating state of the
safety cabinet and automatically disables those functions of the operating panel that are
in contradiction to the safety requirements of the switching state.
The easy operator guidance ensures that all basic functions can be controlled with only a
few operating steps and that the safety cabinet can be shifted into the working mode.
Switch device into work mode:
Keep the
key depressed until ready signal sounds.
Moving the front window up:
Keep rocker switch depressed.
When the front window reaches the working position, the movement stops
automatically. When the movement starts above the working position, the front
window stops at the maximal opening position.
Stopping the upward moving:
Release rocker switch.
Lowering the front window:
Keep lower side of the rocker switch depressed.
Stopping the downward moving:
Release rocker switch.
Switching the device to OFF-Mode:
Keep the
key depressed until ready signal sounds.
NOTE
The front window can only be moved with the rocker switch.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...