| Technical Data
68
|
50161258
Herasafe 2025
Chapter 12
Electrical Data
Herasafe 2025
0.9
1.2
1.5
1.8
Voltage
Rated voltage
V
230V - 1/N/PE AC - 50/60 Hz
Blower voltage
V
48 V / DC
Current
Power consumption
A
7.0
7.0
9.0
9.0
Leakage current
IEC 1010, EN 61010
mA
< 3.5
Motherboard fusing
A
2 x T15A H
Outlet fusing
A
2x T 5 A
On-site fusing
A
Circuit breaker B 16 / Fuse T 16 A
Equipment fusing (230V)
A
2 x T16A H
Power
Power input max.
W
1675
1675
2010
2010
Protection
Protection class
I
Protection type
IP 20
Overvoltage category
(IEC 1010, EN 61010)
II
Contamination degree
(IEC 1010, EN 61010)
2
Connection lines
Connecting lines
cable (2.5 or 5 m/8.2 or 16.4 ft))
Volume, weights and loads
Herasafe 2025
0.9
1.2
1.5
1.8
Volume
Device volume
m
3
/cf
1.1/39
1.4/49
1.7/60
2.1/74
Floorpan
I/cf
30/1
40/1.4
50/1.7
60/2.1
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...