HERAsafe 2025
50161258 |
35
Start-up |
Chapter 5
2. To attach the device frame [5] to the stand [7], install four Allen screws [6] loosely
into the corresponding threaded holes at the underside of the device.
3. Place the safety cabinet onto the stand so that the Allen screws [6] pass through the
corresponding holes [8] of the retaining tabs [10].
4. Push the device frame [5] in the grooves [9] of the retaining tabs [10] all the way to
the stop.
5. Tighten the four Allen screws [6].
Leveling the Device
The cabinet should be levelled only after it has been positioned.
1. Remove transport protection (foil) from the workplate or from the workplate segments.
2. Lift the workplate or the workplate segments and place it/them onto the front and
rear rails in the sample chamber with the wide line of holes facing forward.
3. Device without stand: Place a bubble level onto the workplate and align the
substructure until the bubble level indicates an exactly horizontal position in all
directions.
4. Device with a stand: Place a bubble level onto the workplate and use the four
levellers to effect a level state in all planes.
When adjusting the device stand height, proceed from right to left and from rear to front.
Summary of Contents for HERASAFE 2025
Page 1: ...Valid 30 08 2017 50155204 ...
Page 6: ...Table of Contents 6 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 8: ...Figures 8 50161258 Herasafe 2030i Maxisafe 2030i ...
Page 16: ... Delivery 16 50155215 HERAsafe 2025 Chapter 2 ...
Page 32: ... Unit Description 32 50161258 HERAsafe 2025 Chapter 4 ...
Page 38: ... Start up 38 50161258 HERAsafe 2025 Chapter 5 ...
Page 54: ... Preparation 54 50161258 HERAsafe 2025i Chapter 7 ...
Page 56: ... Shut down 56 50161258 HERAsafe 2025 Chapter 8 ...
Page 62: ... Cleaning and Decontamination 62 50161258 Herasafe 2025 Chapter 9 ...
Page 72: ... Technical Data 72 50161258 Herasafe 2025 Chapter 12 ...
Page 74: ... Device Log 74 50161258 Herasafe 2025 Chapter 13 ...
Page 75: ...Herasafe 2025 50161258 75 Certificate of Decontamination ...
Page 76: ... Certificate of Decontamination 76 50161258 Herasafe 2025 Chapter 14 ...