15 Preparations for transport
(frost protection)
Thermo Scientific
78
50112712_082007_GB
15 Preparations for transport
(frost protection)
♦
Start the Empty Jacket program (P
3
program key).
The water from the double wall is now drained into the disposal system. This
process may take some time, depending on the operating state.
♦
Provide a collecting vessel (container,
bucket). Place the free hose end into the
collecting vessel.
♦
Press the coupling into the coupling socket
(1).
Ö
When the feed water tank has been
drained via the silicone hose:
♦
Press against the outer edge of the
coupling socket (1) with force: The
coupling disengages from the socket.
♦
Clean the removable drip ledge (2).
♦
Remove the residual water from the tank.
♦
Clean the sieve of the feed water tank under running water.
♦
Clean the entire sterilizing chamber as well as the feed water tank
thoroughly with a moist soft cloth.
Draining the vacuum pump
♦
Place the vessel below the screw cap (see picture).
Screw cap
2
1