17 Opening the door without power
Thermo Scientific
82
50112712_082007_GB
17 Opening the door without power
DANGER
Only use the emergency unlocking aid to open the sterilizer
door when the sterilizing chamber is not under pressure.
♦
Check to ensure that the sterilizing chamber is not under pressure.
Slightly jolt the door. If it is possible to move the door back and forth a bit, the
sterilizing chamber is not under pressure.
♦
Remove the cover for the emergency unlocking aid insertion opening.
♦
Insert the emergency unlocking aid in the emergency unlocking
aid insertion opening (see Figure 28).
Figure 28: Emergency unlocking aid insertion opening
♦
Fasten the hook of the emergency unlocking aid to the eye of the lock
barrel (not visible from outside).
♦
Pull the emergency unlocking aid.
The door opens.
♦
Remove the emergency unlocking aid and attach the insertion cover.
When the sterilizer is reconnected to mains, the sterilization phase display may
show an error message.
♦
Proceed as described in Chapter 14, Errors.
DANGER
If a program is interrupted, the load must be re-sterilized for
safety reasons.
Eye
Lock barrel