15 Preparations for transport
(frost protection)
Thermo Scientific
80
50112712_082007_GB
The locking levers of the door latch themselves for transport.
♦
Pull the mains plug.
INFORMATION
For information on the storage conditions, please refer to
Technical Specifications.
INFORMATION
It might be necessary to execute recommissioning according
to valid regulations (German Federal Armed Forces).
15.1 Recommissioning
Every intervention at the sterilizer and / or its supplies, such as changes, rebuilding
and retrofitting that may result in influencing the performance of the sterilizer,
require recommissioning.
When recommissioning, the commissioning tests are checked and, if necessary,
repeated.
Following tests must be executed at commissioning:
•
Concordance with specifications
•
Calibration of the units concerning control, display and documentation
•
Checking measuring, controlling and regulating features
•
Checking operational safety
•
Checking vacuum-tightness
•
Checking consumables
•
Visual inspection