2 Safety considerations
Thermo Scientific
14
50112712_082007_GB
2.1.2 TERMINATION program
DANGER
Should any anomalies and / or problems occur while the
sterilizer is ready for use or while running a sterilization cycle:
Press the green START/STOP key to terminate any ongoing
programs, if necessary.
Examples of possible anomalies or problems:
•
Steam escapes from behind the cover or from the disposal area;
•
Water dropping from the bottom of the unit;
•
Incorrect choice of program with adverse consequences for the items to
be sterilized (e.g. excessive sterilization temperatures).
Pressing the START/STOP key terminates the sterilizing program. The device will
enter a state in which the load can be safely removed.
If you had inadvertently chosen an inappropriate program:
♦
Select the correct sterilizing program for your load.
♦
Start the newly selected sterilizing program.
In case of anomalies and problems:
♦
Contact the authorized service:
DANGER
If a program is interrupted, the load must be re-sterilized for
safety reasons.