Thermo Scientific
Chapter 6
|
Specifications and part numbers
HyPerforma Smartainer 3.0 User’s Guide
|
55
Table 6.7. 3,000 L Smartainer 3.0 unit specifications.
Specification
Jacketed
Non-jacketed
Un
it ge
om
et
ry
Rated liquid working volume
3,000 L
Total volume (liquid and gas)
3,300 L
Interior length
129.54 cm (51 in.)
Interior width
120.65 cm (47.5 in.)
Interior height
212.59 cm (97.7 in.)
Overall length
140.46 cm (55.3 in.)
139.7 cm (55 in.)
Overall width
131.06 cm (51.6 in.)
130.81 cm (51.5 in.)
Overall height
282.57 cm (111.25 in.)
Extra height (for lid)
139.19 cm (54.8 in.)
Te
mp
er
at
ur
e c
on
tr
ol
Jacket area
91,438.84 cm
2
(14,173.05 in
2
)
N/A
Jacket volume
54.4 L
N/A
Maximum jacket pressure
5.17 bar (75 psig)
Jacket flow rate at 3.4 bar (50 psi)
122.5 L/min.
N/A
Process connection
1.5 in. Sanitary tri-clamp
N/A
Recommended TCU model: maximum heating/cooling
TF24000: 22500/24000 W
N/A
Approximate liquid heat-up time, hours:minutes
(5–37°C)—100% volume
6:16 (6:21 with circulation)
N/A
Approximate liquid cool down time, hours:minutes
(37–5°C)—100% volume
13:26 (12:22 with circulation)
N/A
Gen
er
al
Dry skid weight* (mass)
1,160 kg (2,557 lbs)
836 kg (1,844 lbs)
Wet skid weight*—rated working volume (mass)
4,160 kg (9,171 lbs)
3,836 kg (8,458 lbs)
Pneumatic supply requirement, where applicable
Minimum 5 bar (72.5 psi)
Required tubing size for hoist pneumatic air line
8 mm
Load cell display power supply requirement (voltage,
phase, amp)
85–264 VAC, single, 10/20 A
Temperature display power supply requirement (voltage,
phase, amp)
85–264 VAC, single, 10/20 A
Noise level
< 70 dB at 1.5 m
Re
comm
en
de
d
op
er
at
in
g
par
am
et
er
s
Environmental operating temperature range
Ambient to 40 ± 0.5°C (104 ± 0.9°F)
Volume range
0–3,000 L
Maximum BPC pressure
0.03 bar (0.5 psi)
*Vessel-only estimate. Accessories will add additional weight.