2.2.2 Moving the Smartainer 3.0
For convenient transport of the Smartainer 3.0, four casters may be
attached to the base in units 1,000 L and smaller. Use the following
steps to move Smartainer 3.0 units on casters.
1. Ensure that the load cells are locked out (see previous section).
2. Release the caster brakes found on the mounting brackets of the
casters by moving them to the Up (released) position.
3. Carefully apply even pressure to the Smartainer 3.0 to roll it to the
new desired location. Use the handle where applicable.
4. Once in position, ensure that all caster brakes have been reset to
the Down (locked) position.
5. The load cells can now be unlocked.
If lifting is required, use a forklift to move the Smartainer 3.0 unit. See
the Smartainer 3.0 drawings for information on weights in order to
select an appropriate forklift. To lift the Smartainer 3.0, ensure that
the forklift tines are sufficiently long and spaced far enough apart for
stability. Forklift tines should be placed in the forklift hole locations on
the unit (Figure 2.11) and spaced as close to the casters or stationary
feet as possible.
Figure 2.10. Using a wrench to tighten the
bolt.