Thermo Scientific
Chapter 3
|
Operating information
HyPerforma Smartainer 3.0 User’s Guide
|
24
Figure 3.12. Fully inflated BPC inside the unit.
Note:
It is recommended for containers 1,000 L and larger that the
BPC is also inflated to ensure proper BPC and port positioning.
• If equipped, and prior to inflation, the BPC hoist should be
raised to its highest level. The BPC hoist will help hold the BPC
in the correct position during installation.
• Inflate the BPC through one of the top lines of the BPC based
upon available air connections. Do not exceed 25 slpm or
0.034 bar (0.5 psi) internal BPC pressure.
• Inflation time will vary based on flow rate, inlet pressure, and
container volume.
• As the BPC inflates, ensure that the ports are properly oriented
in the container.
WARNING:
The BPC is not rated as a pressure vessel. DO NOT
EXCEED 0.034 bar (0.5 psi) within the BPC or the system could fail,
causing personal injury or damage to equipment. DO NOT leave the
BPC unattended while inflating.
Figure 3.12 shows the inflated BPC inside of the Smartainer unit.