If the Smartainer 3.0 unit needs to be moved, the load cells must be
locked again before transport. To lock load cells for stationary units,
use the following steps.
1. Place the solid steel plate (included in the load cell unlocking
kit, Mettler Toledo part #30219938) in the former location of the
large plastic locking plate (Figure 2.9). To lower the bolt, use an
adjustable wrench to turn the bolt counter-clockwise. This will
close the screw jack.
Figure 2.8. Removing large locking plate (only
required after initial uncrating).
2. Replace the plastic centering spacers with the stainless steel
spacers (included in the load cell unlocking kit, Mettler Toledo part
#30219938). Use an adjustable wrench to tighten the bolt (Figure
2.10). Once it begins to feel tight, tighten another 1/4 turn to ensure
it is fully tightened. Both centering spacers should be below the
bolt when tightened.
Figure 2.9. Replacing red plastic locking plate
with stainless steel plate.