5
Routine and Preventive Maintenance
Maintaining the APCI Probe
70
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Installing the APCI Capillary
Y
To install the APCI capillary into the APCI probe
1. Using tweezers, insert the APCI capillary into the PEEK tube insert and graphite Vespel
ferrule assembly.
2. Using tweezers, insert the APCI capillary, PEEK tube insert, and graphite ferrule
assembly into the probe. See
3. Adjust the position of the APCI capillary:
a. Holding the APCI probe in the vertical position, carefully shake the PEEK tube
insert into the probe until it meets resistance.
b. Using tweezers, push the capillary into the probe until it is flush with the PEEK
sleeve.
c.
Visually check that the capillary is protruding from the tip of the probe.
Figure 51.
Inserting the capillary, PEEK sleeve, and graphite ferrule into the APCI probe
Figure 52.
APCI capillary, PEEK sleeve, and graphite ferrule inserted into probe
Bottom of the probe mount
Graphite Vespel ferrule
PEEK sleeve insert
PEEK sleeve
Graphite Vespel ferrule
Capillary