5
Routine and Preventive Maintenance
Maintaining the ESI Probe
64
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Figure 44.
View of ESI probe mount, showing the insertion of the capillary
3. Adjust the position of the ESI capillary:
a. Holding the ESI probe in the vertical position, carefully shake the PEEK tube insert
into the probe until it meets resistance.
b. Using tweezers, push the capillary into the probe until it is flush with the PEEK
sleeve.
c.
Visually check that the capillary is protruding from the tip of the probe. As
shows, the capillary should protrude from the tip of the probe.
Figure 45.
View of capillary inserted into the ESI probe mount
ESI capillary
Graphite ferrule
PEEK sleeve insert
Internal bottom of
probe mount
ESI probe mount
Capillary
Internal bottom of
ESI probe mount