5
Routine and Preventive Maintenance
Maintaining the Source Block Assembly
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
83
Cleaning the Entrance Cone and the Cone Wash Nozzle
The cone wash nozzle requires cleaning only if it becomes blocked.
Y
To clean the entrance cone and the cone wash nozzle
1. Use a 2.5 mm flat blade screwdriver to remove the O-ring on the back of the entrance
cone. See
. For O-ring part numbers, see
on
Figure 64.
Backside of the entrance cone, showing its O-ring (BS010)
2. Sonicate the entrance cone in 100% methanol. If methanol does not remove the
contamination, sonicate the cone in a 10% v/v aqueous solution of formic acid, rinse
with distilled water, and then rinse with 100% methanol.
3. Using a microscope set to 30x magnification, inspect the inside of the cone to ensure
cleanliness. See
Figure 65.
View of the center of the entrance cone
O-ring, BS010
6.07 mm ID × 1.78 mm THK
Orifice in the center of the entrance cone
CAUTION
Because
solvent and acid can damage them, do not sonicate O-rings.
CAUTION
Take care when handling the cone wash nozzle. Its tip is extremely fragile.