5
Routine and Preventive Maintenance
Maintaining the Source Block Assembly
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
85
11. Lock the probe heater in place by turning its black knobs forward 90 degrees. See
.
12. Reinsert the API probe into the probe heater and reattach the PEEK fingertight fitting to
the probe. See
.
Removing the Source Block Assembly
You must clean the entire source block assembly on a regular basis if you inject complex
sample matrices or use highly buffered mobile phases. To clean the source block assembly, you
must remove it from the instrument and disassemble it.
Y
To remove the source block assembly from the MS detector
1. If you have not already done so, do the following:
• Prepare your LC/MS system for maintenance as described in
.
• Remove the probe and the probe heater as described in
• Remove the entrance cone and the cone wash nozzle as described in
Entrance Cone and the Cone Wash Nozzle”
2. Loosen the thumbscrews on the source block and pull the source block assembly out of
the MS detector. See
.
Figure 67.
View of the source block assembly being removed from the MS detector
CAUTION
Wait for the source block and probe heater assembly to cool before carrying out
any maintenance.
Thumbscrews