Routine and Preventive Maintenance
Maintaining the APCI Probe
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
69
3. Using the probe clamp, unscrew the capillary retaining nut.
4. Carefully shake the graphite vespel ferrule (P/N 6070119), PEEK tube insert
(P/N FM102591) and APCI probe capillary (P/N FM102594) out of the probe.
Figure 50.
View of partially disassembled APCI probe with capillary removed
5. Pull the APCI capillary out of the PEEK tube insert, and then do one of the following:
• If the capillary is partially blocked or blocked, dispose of it.
• If the capillary is reusable, clean its surface with [50:50] methanol / water.
CAUTION
Exercise care when handling the APCI capillary because it is fragile and
can be damaged easily.
Probe clamp
Locking plate
Graphite
ferrule
APCI probe mount
M3 × 10, cap head
stainless steel screws
Capillary
retaining nut
PEEK tube insert
APCI capillary tube