Thermo Scientific
Surveyor MSQ Plus Hardware Manual
119
A
Optimizing the LC Conditions
This appendix includes helpful information for optimizing your LC system and your
Cone Wash system.
Flow Rates
In general, the LC column employed determines the choice of flow rate. Each column has an
optimum flow rate. The guidelines in
The different ionization modes require different flow rates and column IDs. The following
guidelines apply when using the Surveyor MSQ Plus MS detector:
• Electrospray can operate at all the flow rates described in
.
• APCI cannot operate at flow rates below 0.2 mL/min. Therefore, suitable column IDs are
2.1 mm, 3.9 mm and 4.6 mm.
Contents
• LC Solvents and Mobile Phase Additives
•
Table 8.
LC columns and flow rates
Column ID
Flow Rate
4.6 mm
1.0 mL/min
3.9 mm
0.5 mL/min
2.1 mm
0.2 mL/min
1.0 mm
40-50
μ
L/min
Capillary
<10
μ
L/min