4
Switching Probes
Switching From ESI to APCI
52
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
7. Turn the locking plate on the APCI probe clockwise to the open position. Insert the
APCI probe into the probe heater as shown in
. Then, turn the locking plate
counterclockwise to the closed position.
shows the Surveyor MSQ Plus MS detector setup for the APCI mode.
shows the corona pin in the operational position for the APCI mode.
8. Turn the corona pin knob 90 degrees to its horizontal position.
9. Insert the PEEK fingertight fitting into the APCI probe and screw in.
10. Adjust the nitrogen gas pressure to 310 kPa (45 psi).
Figure 32.
Surveyor MSQ Plus MS detector setup for the APCI mode
CAUTION
Take care not to damage the capillary of the probe as you insert the APCI
probe into the probe heater.
Locking plate in
locked position
Corona pin knob
Source housing
Probe mount
Probe heater