Routine and Preventive Maintenance
Maintaining the Source Block Assembly
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
91
Figure 72.
View of the source block with the proper alignment of the extraction cone
insulator, extraction cone, and hexapole insulator
Installing the Source Block Assembly
You must reinstall the source block assembly before you prepare the system for operation.
Y
To install the source block assembly into the MS detector
1. Insert the source block assembly (P/N FM102573) into the source block enclosure, and
then tighten the thumbscrews on the source block. See
2. Handling the cone wash nozzle by its base because the tip is fragile, insert the cone wash
nozzle (P/N FM102521) into the source block, and then carefully rotate the nozzle tip to
the right. See
Properly aligned components
CAUTION
The corona pin is very sharp. Do
not
attempt to install the entrance cone or
the cone wash nozzle before you turn the corona pin knob to its vertical position. See
CAUTION
The tip of the cone wash nozzle is very fragile.