2
Functional Description
MS Detector
24
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Figure 15.
View of the ESI probe, showing the principal components and the pressure regions
LC eluent
Nebulizing gas, N
2
Sheath gas, N
2
Probe heater
Insert capillary
Entrance cone
RF/dc prefilter
RF/df prefilter
vacuum region
M-Path
vacuum region
Atmospheric
pressure region
1 Torr
10
-3
Torr
Forepump