121
Shandon Excelsior ES Operator Guide
Appendix C Program Examples
These are the default programs installed on Excelsior ES.
REAGENT
HOLD / USE
TIME
DRAIN
VACUUM
STIR
1
Formalin
Amb
00:30
30
Off
5
2
Formalin
Amb
00:30
60
Off
5
3
Alcohol
30
01:00
30
On
5
4
Alcohol
30
01:00
30
On
5
5
Alcohol
30
01:00
30
On
5
6
Alcohol
30
01:00
30
On
5
7
Alcohol
30
01:00
30
On
5
8
Alcohol
30
01:00
60
On
5
9
Xylene
30
01:00
30
On
5
10
Xylene
30
01:00
30
On
5
11
Xylene
30
01:00
60
On
5
12
Wax
60
01:20
120
On
5
13
Wax
60
01:20
120
On
5
14
Wax
60
01:20
120
On
5
REAGENT
HOLD / USE
TIME
DRAIN
VACUUM
STIR
1
Formalin
Amb
00:30
30
Off
5
2
Formalin
Amb
00:30
60
Off
5
3
Alcohol
30
01:00
30
On
5
4
Alcohol
30
01:00
30
On
5
5
Alcohol
30
01:00
30
On
5
6
Alcohol
30
01:00
30
On
5
7
Alcohol
30
01:00
30
On
5
8
Alcohol
30
01:00
60
On
5
9
Xylene
30
01:00
30
On
5
10
Xylene
30
01:00
30
On
5
11
Xylene
30
01:00
60
On
5
12
Wax
60
01:20
120
On
5
13
Wax
60
01:20
120
On
5
14
Wax
60
01:20
120
On
5
Introduction
Routine Overnight
Program
Rapid Biopsy
Program
Summary of Contents for Shandon Excelsior ES
Page 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Page 5: ...5 Shandon Excelsior ES Operator Guide...
Page 12: ...12 Shandon Excelsior ES Operator Guide...
Page 22: ...22 Shandon Excelsior ES Operator Guide...
Page 62: ...62 Shandon Excelsior ES Operator Guide...
Page 101: ...101 Shandon Excelsior ES Operator Guide...
Page 103: ...103 Shandon Excelsior ES Operator Guide...
Page 104: ...104 Shandon Excelsior ES Operator Guide...