75
Shandon Excelsior ES Operator Guide
1. Replace the wax.
a. Open the cabinet doors, and take out the waste wax tray
ww.
The Quality Control screen appears.
b. Take the second waste wax tray
ww
supplied with the
instrument and put it, with a new liner, back in the cabinet.
In the Quality Control screen, highlight
ww
and select
[Confirm Empty].
c. Add paraffin wax pellets to the empty wax bath W3.
Highlight wax, and select
[START hEATiNG]
. While the
wax is heating, go to 2 below.
2. Replace alcohol.
a. Take out container
Ex1
(waste alcohol) and dispose of.
b. Place fresh 100 % alcohol into position Ex1.
In the Quality Control screen highlight the text under alcohol
and select
[NEw].
3. Before continuing, ensure that the wax level is correct.
Running your next process completes the reagent rotation.
Empty exchange containers