Shandon Excelsior ES Operator Guide
Ensure that you have all parts on the packing list.
If parts are missing or broken, contact your dealer.
Notes
1. Quote the instrument Serial Number, your Order Number,
Invoice Number, Delivery Note (or Packing Slip) Number and
the date.
2. If you need to transport the instrument, refer to
Appendix D for repacking instructions.
Excelsior ES must be level from front to back. Before loading
with reagents and wax, ensure that it is fully adjusted.
EXCELSIOR ES WEIGHS APPROXIMATELY 112 kg (246 lb)
WHEN EMPTY. WHEN FILLED WITH REAGENTS AND
WAX, Excelsior ES WEIGHS 200 kg (440 lb).
s
Move Excelsior ES to its final position. The floor should be level -
the floor and any covering must be non-flammable. There should
be at least 95 mm (4 inches) behind the instrument.
To ensure this, leave the transit handles fitted.
Open the reaction chamber lid and remove the dessicant pack
from the chamber. Pour 50ml of water into the reaction chamber
and use it as a fluid level. When the instrument is level, the water
settles evenly at the bottom of the chamber.
If necessary, adjust the front castors. Loosen the locknut
(a wrench is provided), then turn the castor with the adjustment
rod provided.
When the instrument is level, tighten the locknut.
Mop up the water and dry the reaction chamber.
Levelling the Instrument
Summary of Contents for Shandon Excelsior ES
Page 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Page 5: ...5 Shandon Excelsior ES Operator Guide...
Page 12: ...12 Shandon Excelsior ES Operator Guide...
Page 22: ...22 Shandon Excelsior ES Operator Guide...
Page 62: ...62 Shandon Excelsior ES Operator Guide...
Page 101: ...101 Shandon Excelsior ES Operator Guide...
Page 103: ...103 Shandon Excelsior ES Operator Guide...
Page 104: ...104 Shandon Excelsior ES Operator Guide...