80
Shandon Excelsior ES Operator Guide
The screen shows events monitored by Excelsior ES, which can
trigger alarms, alerting operators to act, and ensure processing
quality. To change settings,
1. Highlight an event.
2. Select from the sounding options: Use the number keys for
repeated sounds or
[CoNTiNUoUS]
to sound the alarm until it
is silenced with the
button on the keypad.
3. For the remote alarm, highlight under
RA1
and
RA2
and
select either
[off]
or
[oN].
Different tunes can be selected to play for each event. Both
remote alarms can be selected and operated for any combination
of events.
External circuits should only be connected to the Remote Alarm
socket by a technically competent person. The external circuit
should comply with the requirements of IEC1010-1 and or
IEC950.
Excelsior ES can send messages via the serial
D-connector, to report events listed on the LIMS screen.
Use this option to bring any concealed reagent or wax into the
reaction chamber, outside of processing. The Reagent and Wax
Inspection Screen is shown.
Set LIMS
Interface
Reagent and Wax
Checks