73
Shandon Excelsior ES Operator Guide
1. Replace the wax
a. Open the cabinet doors, and take out the waste wax tray
WW. The Quality Control screen appears.
b. Take the second waste wax tray WW supplied with the
instrument, and put it, with a new liner, back in the
cabinet.
In the Quality Control screen, highlight WW and select
[CoNfiRM EMPTy].
c. Add paraffin wax pellets to the empty wax bath W3.
Highlight wax, and select [
START hEATiNG]
. While the
wax is heating, go to 2 below.
2. Replace alcohol and xylene.
a. Take out containers Ex1 (waste alcohol) and Ex2 (waste
xylene) and dispose of them.
b. Place fresh 100 % alcohol and fresh xylene into positions
Ex1
and
Ex2.
In the Quality Control screen highlight the text under alcohol
and select
[NEw
]. Highlight the text under xylene and select
[NEw].
3. Transfer reagents to concealed bottles.
Before continuing, ensure that the wax level is correct.
Running your next process completes the reagent rotation.